1-Hydroxypyrene in Urine | On-Line Analysis
Complete Kit, Order No. 50 000


Urinary 1-hydroxypyrene is used as a biological index for the assessment of an exposure to polycyclic aromatic hydrocarbons (PAH). Many individual PAHs are genotoxic carcinogens and are formed by the incomplete combustion of fossis fuels.

The ClinTox® complete kit of RECIPE is designed for the automated determination of urinary 1-hydroxypyrene in order to estimate an environmental or occupational related PAH exposure. The sample preparation is performed on-line by column switching technique using a highly selective SPE-column (SPE: solid phase extraction).

 
  • Fully automated method

  • Recovery 100 %

  • High analytical sensitivity


Test Data

 

HPLC Parameters

Linearity: 100 - 24000 ng/l
Recovery: 100 %
Lower detection limit: 20 ng/l
Lower determination limit: 100 ng/l
Intraassay precision: 2.6 %
Interassay precision: 2.9 %
 
Pump 1: gradient pump,
flow rates: 1.0, 1.5 ml/min
Pump 2: isocratic pump:
flow rate: 1.0 ml/min
Injection volume: 250 µl
Injection interval: 36 min
Fluorescence detector: 242 nm (exc.), 388 nm (em.)
HPLC-Thermostat: 35°C

 

Sample Preparation

50000-flow

 

Chromatograms

ClinChek® Urine Control, Level I ClinCheck® Urine Control, Level II
50000_chrom1   50000-chrom2
1-Hydroxypyrene: 18.17 min   1-Hydroxypyrene: 18.00 min


1-Hydroxypyrene in Urine | On-Line Analysis
Complete Kit, Order No. 50 000

Ordering Information

Order No. Description Quantity
50000  ClinTox® Complete Kit for 1-Hydroxypyrene in Urine for 100 assays 1 pce. 
 
  Kit Components (order no. 50000):  
50007 Mobile Phase A 1500 ml
50008 Mobile Phase B 000 ml
within the kit: 2 x order no. 50008
50009 SPE - Buffer 2500 ml
within the kit: 2 x order no. 50009
9969 Urine Calibrator, lyophil. 5 x 10 ml
50021 S Stabilizing Reagent 200 ml
50022 Enzyme Solution 1 ml
  Manual 1 pce.
  Quick Reference 1 pce.
 
  Start Accessories:  
50030 Analytical Column with test chromatogram 1 pce.
FK7000 manu-CART® cartridge holder, one time accessory 1 set
50031 SPE - Column 1 pce.
50032 Guard Column 5 pcs.
  
  ClinChek® Controls for toxic organic compounds:  
8867 Urine Control, lyophil., Level I 10 x 10 ml
8868 Urine Control, lyophil., Level II 10 x 10 ml
8869 Urine Control, lyophil., Level I, II 2 x 5 x 10 ml


back to the top    back to the top back to the top back to ClinTox Human Biomonitoring Overview